Comparison of Recordings from Microelectrode Arrays and Single Electrodes in the Visual Cortex
Ryan C.Kelly, Matthew A.Smith, Jason M.Samonds, Adam Kohn, A.B.Bonds, J.Anthony Movshon, Tai Sing Lee | The Journal of Neuroscience | 2007 | 10.1523/JNEUROSCI.4906-06.2007 | PDFHere we focus on a specific microelectrode device, the Cyberkinetics “Utah” Array (Cyberkinetics Neurotechnology Systems, Foxborough, MA) (Fig. 1A). This device is a 10 × 10 grid of silicon microelectrodes (1 mm in length) spaced 400 μm apart, covering 12.96 mm2. We analyzed recordings from single electrodes and the Utah array in macaques and cats. Previous reports on the quality (Nordhausen et al., 1996) and long-term stability (Suner et al., 2005) of array recordings have not made a quantitative, direct comparison with established single-electrode recordings. In light of the increasing popularity of the array, we addressed this uncertainty by comparing waveforms recorded with the array to waveforms recorded with accepted single-electrode techniques. We found that the array yields good recordings on a large number of electrodes, with qualities comparable to those from single-electrode recordings. On average, the recording quality is somewhat lower than that of single electrodes but, nonetheless, is sufficient for assessing tuning properties such as the spatiotemporal receptive field (STRF) and orientation tuning.